Arca
Skip to content
Tools
Tools

Ariatele Telemetry Transmitter (96281)

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K121480, from Spacelabs Healthcare (Nederland, CO). Product code MHX, class 2.

As of Nov 21, 2026, K121480 is an FDA 510(k) held by Spacelabs Healthcare, substantially equivalent on Jun 6, 2012 after 19 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 19 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

K Spacelabs Healthcare AriaTele Telemetry Transmitter (96281) The Spacelabs Healthcare AriaTele Telemetry Transmitter (96281), when used in conjunction with a Spacelabs Healthcare Ultraview patient monitor and telemetry receiver, provides a means for the continuous monitoring of electrocardiographic signals in order to detect abnormal cardiac rhythms, including life-threatening events such as…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K121480, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 2, 2014Z-2586-2014The transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions. This is an expansion of recall Z-0871-2014…Terminated
Dec 19, 2013Z-0871-2014The AriaTele Model 96281 with SpO2 monitoring Option C is recalled because the AriaTele (transmitter) display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the…Terminated

Other MHX clearances

The newest 510(k)s for the same product code; their median review was 207 days; see every MHX device

Record

Type
Special
Received
May 18, 2012
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files