Arca
Skip to content
Tools
Tools

Software Syngo Mr D13A for The Magnetom Systems Aera/Skyra/Avanto/Verio

System, Nuclear Magnetic Resonance Imaging. 510(k) K121434, from Siemens Medi Cal Solutions, Inc. (Malvern, PA). Product code LNH, class 2.

As of Nov 21, 2026, K121434 is an FDA 510(k) held by Siemens Medi Cal Solutions, Inc., substantially equivalent on Nov 5, 2012 after 175 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 175 days
  • Recalls: 3
  • Cited as a predicate by: 17 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

evice Names: Software syngo MR D13A for the MAGNETOM systems Aera/Skyra/Avanto/Verio

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

17 later 510(k) summaries name this device as a predicate; the 12 newest

Recalls

Recalls FDA ties to K121434, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 28, 2025Z-2666-2025, Z-2665-2025There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a…Open
Mar 9, 2015Z-1308-2015The gradient output supervision was permanently turned off on the MAGNETOM system, meaning that gradient outputs could exceed IEC60601-2-33 limits and peripheral nerve stimulation could occur.Terminated

Other LNH clearances

The newest 510(k)s for the same product code; their median review was 143 days; see every LNH device

Record

Type
Traditional
Received
May 14, 2012
Review panel
Radiology
Regulation
21 CFR 892.1000

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.