Accelerator, Linear, Medical. 510(k) K121328, from Elekta , Ltd. (Crawley, West Sussex, GB). Product code IYE, class 2.

As of Nov 20, 2026, K121328 is an FDA 510(k) held by Elekta , Ltd., substantially equivalent on Jun 12, 2012 after 40 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 40 days
  • Recalls: 3
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed. The associated Integrity R3.0 software is the interface and control software for the Elekta medical digital linear accelerator and is intended to assist a licensed practitioner in the delivery of radiation to defined target volumes (e.g.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • MLCI2K082122 · Elekta , Ltd. · Aug 29, 2008

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • AgilityK123808 · Elekta , Ltd. · Apr 4, 2013

Recalls

Recalls FDA ties to K121328, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 25, 2013Z-1084-2014During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.Terminated
Apr 30, 2013Z-1402-2013Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.Terminated
Jul 27, 2012Z-2340-2012Information released with the product Agility did not originally contain information on compatibility with the add-on Dynamic Multileaf Collimator Apex.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
May 3, 2012
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files