Spine Frontier Indus Acp System
Appliance, Fixation, Spinal Intervertebral Body. 510(k) K121060, from Spinefrontier, Inc. (Beverly, MA). Product code KWQ, class 2.
As of Nov 17, 2026, K121060 is an FDA 510(k) held by Spinefrontier, Inc., substantially equivalent on Jul 3, 2012 after 88 days, with 6 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 3, 2012
- Review time: 88 days
- Recalls: 6
- Cited as a predicate by: 3 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFThe SpineFrontier Indus™ ACP System (Indus™ Invue™, Indus™ Invue™ 2, Indus™ Inset™) is intended for anterior spine fixation for use in providing temporary stabilization during the development of cervical spinal fusions. The levels of treatment range from C2 to T1.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Indus Anterior Cervical Plate SystemK093776 · Spinefrontier, Inc. · Mar 26, 2010
- Verteview Anterior Cervical Plate SystemK050588 · Innovative Spinal Design, LLC · Apr 21, 2005
- Nuvasive Helix Mini Acp SystemK073275 · Nuvasive, Inc. · Jan 29, 2008
- Synthes Anterior Cslp SystemK030866 · Synthes (Usa) · Apr 18, 2003
Cited as a Predicate
3 later 510(k) summaries name this device as a predicate- Sacrix® Sacroiliac Joint Fusion Device System, Inspan® ScrewLES Fusion System, Invue®…K232605 · Lesspine Innovations · Sep 28, 2023
- Pyrenees Cervical Plate SystemK153526 · K2m · Apr 27, 2016
- Aspiron Acp SystemK140234 · U&I Corp. · Jun 2, 2014
Recalls
Recalls FDA ties to K121060, newest first; each opens FDA's recall record| Nov 20, 2013 | Z-0574-2014Mismarked and unmarked screws | Terminated |
| Nov 20, 2013 | Z-0573-2014Mismarked and unmarked screws | Terminated |
| Nov 20, 2013 | Z-0576-2014Mismarked and unmarked screws | Terminated |
| Nov 20, 2013 | Z-0575-2014Mismarked and unmarked screws | Terminated |
| Jun 21, 2013 | Z-1933-2013Difficulty removing screws from the Invue caddies. | Terminated |
| May 9, 2012 | Z-1957-2013The nipple of the tipped driver broke off when surgeons were attempting to insert the screw. | Terminated |
Other KWQ Clearances
The newest 510(k)s for the same product code; their median review was 86 days; see every KWQ device- corra® cervical plating systemK262149 · Carlsmed, Inc. · Sep 24, 2026
- Palisades Buttress Plate SystemK261309 · Bethesda Medical Devices, LLC · Jul 16, 2026
- ARIES Anterior Cervical Plate SystemsK252776 · ZheJiang Decans Medical Devices Co., Ltd. · May 14, 2026
- CastleLoc-P Anterior Cervical Plate SystemK261112 · L&K BIOMED Co., Ltd. · May 6, 2026
- corra cervical plating systemK260570 · Carlsmed, Inc. · Apr 22, 2026
- VyPlate Anterior Cervical Plate SystemK260697 · Vy Spine, LLC · Mar 27, 2026
- ANTERIS Thoracolumbar Plate SystemK260015 · SpineCraft · Mar 4, 2026
- PathLoc Lumbar Plate SystemK251940 · L&K BIOMED Co., Ltd. · Mar 3, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Apr 6, 2012
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3060
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files