Alere Cholestech Ldx Analyzer, Alere Cholestech Ldx Lipid Profile - Glu Cassette

Glucose Oxidase, Glucose. 510(k) K120615, from Alere San Diego Inc. Dba Biosite,Innovacon,Hemosen (San Diego, CA). Product code CGA, class 2.

As of Nov 28, 2026, K120615 is an FDA 510(k) held by Alere San Diego Inc. Dba Biosite,Innovacon,Hemosen, substantially equivalent on May 15, 2013 after 441 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 441 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Alere Cholestech LDX® System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile•GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood.

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K120615, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 1, 2024Z-1702-2024There is a potential that the Cholestech LDX battery (an external battery pack available as an optional accessory to the Cholestech LDX Analyzer) may swell which could cause minor burns, cuts or abrasions, or low…Open
Jun 24, 2022Z-1508-2022Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6…Open
Feb 15, 2019Z-2475-2019The polarity of the output of power supply was reversed. Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.Terminated
Aug 2, 2016Z-2778-2016Incorrect expiration date on the magnetic strip that does not match the labeled expiration date.Terminated
Oct 23, 2013Z-0379-2014Alere initiated a recall of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 320167, becuase an incorrect lot number was printed on the outer 10 pack kit box label. The incorrect Lot #…Terminated

Other CGA clearances

The newest 510(k)s for the same product code; their median review was 270 days; see every CGA device

Record

Type
Traditional
Received
Feb 29, 2012
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1345

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files