Arca
Skip to content
Tools
Tools

Macrolyte Dispersive Electrode

Electrosurgical, Cutting & Coagulation & Accessories. 510(k) K120476, from Conmed Corporation (Centennial, CO). Product code GEI, class 2.

As of Nov 19, 2026, K120476 is an FDA 510(k) held by Conmed Corporation, substantially equivalent on Jun 12, 2012 after 117 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 117 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

CONMED® Macrolyte® Dispersive Electrodes are designed for use with electrosurgical generators during electrosurgery and provide a path for, RF energy produced at the active electrode to return to the generator.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K120476, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 19, 2019Z-2545-2019Discrepancy between the correct expiration dates on the individual pouches inside the boxes and incorrect expiration dates on the outer boxes, where the month and day were reversed.Terminated
May 8, 2014Z-1719-2014ConMed Corporation has received complaints regarding sparking, no output and burning at the cord set connection. The complaints have been confirmed by ConMed.Terminated

Other GEI clearances

The newest 510(k)s for the same product code; their median review was 167 days; see every GEI device

Record

Type
Special
Received
Feb 16, 2012
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.4400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files