Infinity Acute Care System, Infinity McAble- Masimo Rainbow Set
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K113798, from Draeger Medical Systems, Inc. (Andover, MA). Product code MHX, class 2.
As of Nov 20, 2026, K113798 is an FDA 510(k) held by Draeger Medical Systems, Inc., substantially equivalent on Mar 22, 2012 after 90 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 90 days
- Recalls: 8
- Cited as a predicate by: 3 devices
The M540 monitors the following parameters: — Heart rate → Arrhythmia (adult and pediatric only) – 12-lead analysis - ST segment analysis including TruST® (adult and pediatric only) - 12-lead ST segment analysis (adult and pediatric only) - Apnea -- Respiration rate – Invasive pressure – Non-invasive pressure - Temperature - Cardiac output (only available when the M540 is docked in an IACS…
- Infinity Acute Care System (Iacs) Monitoring SolutionK093788 · Draeger Medical Systems, Inc. · Mar 8, 2010
- Infinity Masimo Set SP02 PodK061329 · Draeger Medical Systems, Inc. · Jul 7, 2006
- Current Health SystemK231506 · Current Health , Ltd. · Aug 24, 2023
- Infinity Acute Care System (IACS) Monitoring SystemK203088 · Draeger Medical Systems, Inc. · Feb 1, 2022
- Infinity Acute Care System (IACS) Monitoring SolutionK201764 · Draeger Medical Systems, Inc. · Dec 9, 2021
| Dec 23, 2020 | Z-1104-2021The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate… | Open |
| Sep 13, 2019 | Z-0258-2020Cybersecurity vulnerabilities may cause device to reboot, lose alarm functionality, and/or lose communication with cockpit and/or the Infinity Network. | Open |
| Mar 28, 2018 | Z-1813-2018Software anomaly resulting in the loss of patient settings and stored patient data. | Terminated |
| Feb 21, 2018 | Z-1382-2018The invasive pressure visual and audible alarms may not activate. | Terminated |
| Oct 4, 2017 | Z-0942-2018Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Drger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an… | Terminated |
| Jun 6, 2017 | Z-2734-2017Cockpits with revision index 06 or higher that contain 4GB RAM modules may not annunciate audio or visual alarms on the Cockpit and Central Station. | Terminated |
| Dec 10, 2013 | Z-0714-2014After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased… | Terminated |
| Dec 13, 2012 | Z-0939-2013The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the… | Terminated |
- 12SL PrecisionK261307 · Ge Medical Systems Information Technologies, Inc. · Nov 3, 2026
- Radius VSM and AccessoriesK260676 · Masimo Corporation · Oct 14, 2026
- Carescape ONE; Carescape Dock F0; TruSignal ECG MM; TruSignal PRES MM; TruSignal TEMP MM; TruSignal SpO2 MM 1.8M;…K260656 · GE Healthcare Finland Oy · Sep 1, 2026
- Carescape Canvas 1000; Carescape Canvas Smart DisplayK260343 · GE Healthcare Finland Oy · Aug 28, 2026
- Infinity® Acute Care System (IACS) – VG8.1K260579 · Draeger Medical Systems, Inc. · Aug 20, 2026
- Telemetry Monitor (iT30, iT50)K251508 · Edan Instruments, Inc. · Aug 6, 2026
- Monitor B105M; Monitor B125M; Monitor B155M; Monitor B105P; Monitor B125PK253277 · Ge Medical Systems Information Technologies, Inc. · Jul 24, 2026
- Radius VSM and AccessoriesK254209 · Masimo Corporation · Jun 4, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 23, 2011
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files