Infinity Acute Care System, Infinity McAble- Masimo Rainbow Set

Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms). 510(k) K113798, from Draeger Medical Systems, Inc. (Andover, MA). Product code MHX, class 2.

As of Nov 20, 2026, K113798 is an FDA 510(k) held by Draeger Medical Systems, Inc., substantially equivalent on Mar 22, 2012 after 90 days, with 8 recalls tied to it in FDA's recall database.

  • Review time: 90 days
  • Recalls: 8
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The M540 monitors the following parameters: — Heart rate → Arrhythmia (adult and pediatric only) – 12-lead analysis - ST segment analysis including TruST® (adult and pediatric only) - 12-lead ST segment analysis (adult and pediatric only) - Apnea -- Respiration rate – Invasive pressure – Non-invasive pressure - Temperature - Cardiac output (only available when the M540 is docked in an IACS…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K113798, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Dec 23, 2020Z-1104-2021The Infinity M540 patient monitor may randomly reboot due to an error to correctly transmit and read the header data of files in the memory of the device. Under this situation, the device will try to reboot to mitigate…Open
Sep 13, 2019Z-0258-2020Cybersecurity vulnerabilities may cause device to reboot, lose alarm functionality, and/or lose communication with cockpit and/or the Infinity Network.Open
Mar 28, 2018Z-1813-2018Software anomaly resulting in the loss of patient settings and stored patient data.Terminated
Feb 21, 2018Z-1382-2018The invasive pressure visual and audible alarms may not activate.Terminated
Oct 4, 2017Z-0942-2018Dual Hemo MCable Pods with Revision Index (RI) 15 and 16, which are used with the Drger Infinity Acute Care System, may permit liquid ingress inconsistent with their IPX4 rating, which could potentially result in an…Terminated
Jun 6, 2017Z-2734-2017Cockpits with revision index 06 or higher that contain 4GB RAM modules may not annunciate audio or visual alarms on the Cockpit and Central Station.Terminated
Dec 10, 2013Z-0714-2014After a test in which a low SpO2 (yellow) medium grade alarm initiated in neo-natal mode and was subsequently paused, the high grade life threatening SpO2 alarm (red) did not activate when the SpO2 values were decreased…Terminated
Dec 13, 2012Z-0939-2013The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the…Terminated

Other MHX clearances

The newest 510(k)s for the same product code; their median review was 207 days; see every MHX device

Record

Type
Traditional
Received
Dec 23, 2011
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files