Arca
Skip to content
Tools
Tools

Signal Gear Urethral Catheter Electrode

Cystometric Gas (Carbon-Dioxide) On Hydraulic Device. 510(k) K113771, from Neurovision Medical Products, Inc. (Ventura, CA). Product code FAP, class 2.

As of Nov 26, 2026, K113771 is an FDA 510(k) held by Neurovision Medical Products, Inc., substantially equivalent on Dec 13, 2012 after 357 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 357 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Signal Gear Urethral Catheter Electrode is intended for mucosal surface stimulation/recording from the external urinary sphincter for use in conjunction with urodynamic evaluation of the patient. This device is indicated for use in adults, as well as in pediatric patients ages 2 and older.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K113771, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Apr 8, 2021Z-1968-2021, Z-1970-2021The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.Terminated

Record

Type
Traditional
Received
Dec 22, 2011
Review panel
Gastroenterology, Urology
Regulation
21 CFR 876.1620

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc