Arca
Skip to content
Tools
Tools

Imris Mr/X-Ray Head Fixation Device (Hfd)

Holder, Head, Neurosurgical (Skull Clamp). 510(k) K113748, from Imris, Inc. (Winnipeg, Mb, CA). Product code HBL, class 2.

As of Nov 20, 2026, K113748 is an FDA 510(k) held by Imris, Inc., substantially equivalent on Mar 23, 2012 after 93 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 93 days
  • Recalls: 1

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The IMRIS MR/X-ray Head Fixation Device System is an MR safe mechanical support system which is used in head, neck and spine surgery when rigid fixation is required for cranial stabilization. The MR/X-ray HFD is indicated for use during utilization of imaging modalities such as intraoperative MRI, CT imaging, and C-Arm X-ray angiography.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K113748, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jun 26, 2019Z-0163-2020Breaking of the HFD200 Drive Knob force indicator due to over-torqueing of the drive knob beyond the maximum allowable pinning force specification may result in the clinical staff unknowingly pinning the patient's skull…Terminated

Other HBL clearances

The newest 510(k)s for the same product code; their median review was 119 days; see every HBL device

Record

Type
Traditional
Received
Dec 21, 2011
Review panel
Neurology
Regulation
21 CFR 882.4460

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.