Hoffmann Lrf (Limb Reconstruction Frame) System

Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component. 510(k) K113327, from Stryker Corp. (Mahwah, NJ). Product code KTT, class 2.

As of Nov 18, 2026, K113327 is an FDA 510(k) held by Stryker Corp., substantially equivalent on Mar 29, 2012 after 140 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 140 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Stryker Hoffmann LRF (Limb Reconstruction Frame) System is indicated in pediatric patients and adults for the treatment and fixation of: • Open and Closed Fractures - Post-traumatic joint contracture which has resulted in loss of range of motion - Fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction •…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K113327, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 12, 2018Z-1526-2018The wrong parts were picked for packaging. The connecting Nuts were packed in the packaging labeled with Hinge couplingTerminated
Oct 25, 2012Z-0695-2013Two complaints have been filed where Hoffman LRF Wires have broken during load bearing application by patients.Terminated
Jul 3, 2012Z-0266-2013Three complaints have been filed where the Hoffman LRF Telescopic Struts have broken during load bearing application by patients having a body weight greater than 250 LBS.Terminated

Other KTT clearances

The newest 510(k)s for the same product code; their median review was 84 days; see every KTT device

Record

Type
Traditional
Received
Nov 10, 2011
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files