Arca
Skip to content
Tools
Tools

Glucose Test System

System, Test, Blood Glucose, Over The Counter. 510(k) K113137, from Roche Diagnostics (Indianapolis, IN). Product code NBW, class 2.

As of Nov 19, 2026, K113137 is an FDA 510(k) held by Roche Diagnostics, substantially equivalent on Jan 6, 2012 after 74 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 74 days
  • Recalls: 3

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The ACCU-CHEK Nano SmartView Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips or palm. The ACCU-CHEK Nano SmartView Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K113137, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 28, 2021Z-2353-2021Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to…Open
Mar 16, 2021Z-1942-2021Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.Open
Mar 12, 2012Z-1386-2012Accu-Chek Aviva Nano Blood Glucose meters that were manufactured for the purpose of validating a manufacturing process for product to be distributed outside the US were distributed in the US through eBay by an…Terminated

Other NBW clearances

The newest 510(k)s for the same product code; their median review was 213 days; see every NBW device

Record

Type
Special
Received
Oct 24, 2011
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1345

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.