Intellivue MX40 Patient Monitor

Detector And Alarm, Arrhythmia. 510(k) K113125, from Philips Medical Systems (Andover, MA). Product code DSI, class 2.

As of Nov 17, 2026, K113125 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Nov 17, 2011 after 24 days, with 16 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Nov 17, 2011
  • Review time: 24 days
  • Recalls: 16
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K113125, newest first; each opens FDA's recall record
Apr 19, 2023Z-0035-2025During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC…Open, Classified
Feb 2, 2021Z-1671-2021Manufacturing failed to follow process steps resulting in the distribution of potentially defective devicesTerminated
Mar 28, 2019Z-1693-2019The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .Open, Classified
Mar 28, 2019Z-1695-2019The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .Open, Classified
Mar 28, 2019Z-1692-2019The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .Open, Classified
Mar 28, 2019Z-1694-2019The MX40 may experience increased power consumption and may have a lack of visual and/or audible notifications while in a Low Battery State .Open, Classified
Oct 10, 2018Z-0592-2019A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above…Open, Classified
Apr 30, 2018Z-2141-2018Five warning statements are missing from the instructions for use.Terminated
Apr 30, 2018Z-2143-2018Five warning statements are missing from the instructions for use.Terminated
Apr 30, 2018Z-2144-2018Five warning statements are missing from the instructions for use.Terminated
Apr 30, 2018Z-2142-2018Five warning statements are missing from the instructions for use.Terminated
Mar 6, 2017Z-1707-2017Philips lntelliVue MX4O WLAN Patient Wearable Monitor may not automatically switch to Monitor Mode with audible alarms when association with central monitoring system is unsuccessful [incomplete]Terminated
Sep 19, 2016Z-0291-2017Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iXTerminated
Sep 19, 2016Z-0292-2017Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iXTerminated
Sep 19, 2016Z-0293-2017Philips IntelliVue MX40 Patient Wearable Monitor Configuration Setting Disables Generation and Delivery of ECG Alarms to PIIC iXTerminated
May 22, 2014Z-1746-2014IntelliVue MX40 (part numbers 865350, 865352) with software revisions B.00 or B.01 may not resume alarming when the alarms are paused and the pause timer ends.Terminated

Other DSI Clearances

The newest 510(k)s for the same product code; their median review was 248 days; see every DSI device

Record

Type
Special
Received
Oct 24, 2011
Review panel
Cardiovascular
Regulation
21 CFR 870.1025

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files