Clearlink Luer Activated Valve, Clearlink System Non-Dehp Catheter Extension Sets
Set, Administration, Intravascular. 510(k) K112893, from Baxter Healthcare Corporation (Mcgaw Park, IL). Product code FPA, class 2.
As of Nov 20, 2026, K112893 is an FDA 510(k) held by Baxter Healthcare Corporation, substantially equivalent on Oct 18, 2011 after 15 days, with 8 recalls tied to it in FDA's recall database.
- Review time: 15 days
- Recalls: 8
- Cited as a predicate by: 5 devices
For use with a vascular access device for the administration of drugs and solutions. The Clearlink Luer Activated Valve is an in-line injection site, which can be connected to standard male Luer adapters (e.g., syringes or sets) for continuous or intermittent fluid administration or the withdrawal of fluids.
- K003255K003255
- PCA Syringe SetsK212262 · Baxter Healthcare Corporation · Oct 3, 2022
- Intravascular Administration SetsK203609 · Baxter Healthcare Corporation · Sep 30, 2021
- Intravascular Extension Sets and AccessoriesK192366 · Baxter Healthcare Corporation · Jul 20, 2020
- Clearlink System Solution SetK180739 · Baxter Healthcare Corporation · May 28, 2019
- Padi-LockK141200 · Padi-Lock, LLC · Oct 2, 2014
| Aug 29, 2025 | Z-0197-2026, Z-0199-2026, Z-0191-2026, Z-0190-2026, Z-0198-2026IV sets may leak. | Open |
| Aug 9, 2022 | Z-1653-2022Firm noted an increase in customer reports of leaks. | Terminated |
| Sep 4, 2013 | Z-0099-2014The firm is issuing a voluntary recall due to the potential for pinholes in the packaging. Pinholes are a breach in the sterile barrier and could lead to potential contamination. | Terminated |
| Mar 28, 2013 | Z-1271-2013Customers have reported that when separating an individual package from its attached grouping, the adjacent package has opened, compromising its sterile barrier properties. | Terminated |
- Gravity Blood Sets with 200µm FilterK260261 · B.Braun Medical, Inc. · Oct 22, 2026
- Ecomed Disposable Administration SetsK254123 · Ecomed Solutions, LLC · Sep 11, 2026
- iiSure Infusion SetK261530 · Deka Research and Development · Jun 3, 2026
- AMSafe Administration Set (BA-70075, BA-70071-D, BA-70070, BA-70071-P, BA-70072, BA-70072-F, BA-70071-DF,…K252543 · Amsino International, Inc. · May 4, 2026
- PuraCath Firefly Needleless Connector IT (9005)K251375 · Puracath Medical, Inc. · Feb 12, 2026
- nSet+ Stabilization Set [6426733]; nSet+ Stabilization Set with nSyte Needle Free Connector (NFC) [6426731]; nSyte…K251257 · Np Medical · Dec 4, 2025
- BD Alaris Pump Infusion SetK250325 · Carefusion (Bd) · Oct 30, 2025
- EZ™ IV Administration SetK251814 · Epic Medical Pte. , Ltd. · Aug 29, 2025
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Oct 3, 2011
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5440
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files