Antibodies, Anti-Cyclic Citrullinated Peptide (Ccp). 510(k) K112810, from Bio-Rad Laboratories (Benicia, CA). Product code NHX, class 2.

As of Nov 26, 2026, K112810 is an FDA 510(k) held by Bio-Rad Laboratories, substantially equivalent on Oct 26, 2011 after 29 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 29 days
  • Recalls: 2

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The BioPlex® 2200 Anti-CCP kit is a multiplex flow immunoassay intended for the semi-quantitative detection of IgG antibodies to Cyclic Citrullinated Peptide (CCP) in human serum or plasma (EDTA and sodium heparin). Detection of CCP antibodies is used as an aid in the diagnosis of rheumatoid arthritis and should be used in conjunction with other clinical findings and laboratory results.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K112810, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Feb 21, 2023Z-1377-2023, Z-1378-2023Bio-Rad received four customer complaints from 11/30/2022 to 12/21/2022 on BioPlex 2200 Anti-CCP Kit (Lot 301481) stating that results were discrepant while performing lot-to-lot validation against previous lots.…Open

Record

Type
Special
Received
Sep 27, 2011
Review panel
Immunology
Regulation
21 CFR 866.5775

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files