Arca
Skip to content
Tools
Tools

Paltop Dental Implant System

Implant, Endosseous, Root-Form. 510(k) K112795, from Paltop Advanced Dental Solutions, Ltd. (Caesarea, IL). Product code DZE, class 2.

As of Nov 18, 2026, K112795 is an FDA 510(k) held by Paltop Advanced Dental Solutions, Ltd., substantially equivalent on Apr 20, 2012 after 207 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 207 days
  • Recalls: 5
  • Cited as a predicate by: 9 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Paltop Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

9 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K112795, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 30, 2019Z-0019-2020Surface micro/nano topography on some of the implants contained within this lot may not be consistently present over the entire surface of the implant.Terminated
Feb 22, 2019Z-1063-2019, Z-1062-2019, Z-1061-2019, Z-1064-2019Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw featureTerminated

Other DZE clearances

The newest 510(k)s for the same product code; their median review was 257 days; see every DZE device

Record

Type
Traditional
Received
Sep 26, 2011
Review panel
Dental
Regulation
21 CFR 872.3640

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.