Arca
Skip to content
Tools
Tools

Restoris Partial Knee Application (Pka), Rio Standard System

Orthopedic Stereotaxic Instrument. 510(k) K112507, from Mako Surgical Corporation (Fort Lauderdale, FL). Product code OLO, class 2.

As of Nov 19, 2026, K112507 is an FDA 510(k) held by Mako Surgical Corporation, substantially equivalent on Mar 1, 2012 after 184 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 184 days
  • Recalls: 6
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Restoris Partial Knee Application (PKA), for use with the Robotic Arm Interactive Orthopedic System (RIO), is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K112507, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 24, 2018Z-2924-2018Specific lots of the Vizadisc Knee Procedure Tracking Kit and Vizadisc Hip Procedure Tracking Kit have the potential to be damaged on the Vizadisc reflective material causing an inability to be detected by the camera.Terminated
Aug 7, 2017Z-3130-2017Software discrepancy of not showing all the EE constants, when the screen is filled.Terminated
Jun 7, 2017Z-2741-2017An intermittent electrical problem that could lead to a loss of system power due to a loose screw connection.Terminated
Jul 16, 2015Z-0224-2016The locking pin on the RIO Base Array is damaged during assembly onto the RIO Base Array Connector.Terminated
Jan 22, 2015Z-1128-2015When using the MAKOplasty partial knee Arthroplasty application, the burr continues spinning outside of the stereotactic boundary and after the control switches (foot pedal and trigger) cease to be activatedTerminated
Apr 30, 2014Z-1508-2016Revision of the field manualTerminated

Other OLO clearances

The newest 510(k)s for the same product code; their median review was 91 days; see every OLO device

Record

Type
Traditional
Received
Aug 30, 2011
Review panel
Neurology
Regulation
21 CFR 882.4560

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.