Tools

Distal Volar Radius Plating Systems

Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite. 510(k) K112345, from Medos International SARL (Warsaw, IN). Product code LXT, class 2.

As of Nov 18, 2026, K112345 is an FDA 510(k) held by Medos International SARL, substantially equivalent on Oct 31, 2011 after 77 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 31, 2011
  • Review time: 77 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K112345, newest first; each opens FDA's recall record
Nov 20, 2018Z-0708-2019One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screwsTerminated
Nov 20, 2018Z-0709-2019One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screwsTerminated
Jan 29, 2016Z-1082-2016Components of the DVR ePAKs can oxidize during shipment and storage prior to use. A delay in surgery, less than 30 minutes, may occur if the surgical staff proceeds to re-sterilize the product. If oxidative material…Terminated
Mar 18, 2015Z-1446-2015A Biomet investigation found that the Low Profile Non-Locking Screw 2.7mmx22mm screw (Part Number: 131227222 Lot: RM105E) was anodized green and should have been anodized gold per the print.Terminated

Other LXT Clearances

The newest 510(k)s for the same product code; see every LXT device

Record

Type
Traditional
Received
Aug 15, 2011
Review panel
Orthopedic
Regulation
21 CFR 888.3030

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files