Viewray System
Accelerator, Linear, Medical. 510(k) K111862, from Viewray, Incorporated (Oakwood Village, OH). Product code IYE, class 2.
As of Nov 20, 2026, K111862 is an FDA 510(k) held by Viewray, Incorporated, substantially equivalent on May 22, 2012 after 327 days, with 9 recalls tied to it in FDA's recall database.
- Review time: 327 days
- Recalls: 9
- Cited as a predicate by: 1 device
The ViewRay System, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
- Trilogy MXK092871 · Varian Medical Systems, Inc. · Nov 30, 2009
- The ViewRay (MRIdian) Linac SystemK162393 · Viewray, Incorporated · Feb 24, 2017
| Oct 20, 2020 | Z-0481-2021There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user. | Terminated |
| Nov 18, 2019 | Z-1803-2020Registering dose and structures in the treatment delivery workflow could result in an alignment discrepancy between the imported previously delivered dose and the displayed patient anatomy and structures. The MRIdian… | Terminated |
| Mar 3, 2019 | Z-0676-2020Issue with transferring treatment regiment from one system to another. Adequate instructions for transferring treatment plan(s) are not provided within the operator's manual or within the software operating system. | Terminated |
| Mar 7, 2016 | Z-1309-2016When editing the isocenter or the couch position of the plan while in the treatment workflow (in the Points screen) and re-optimizing, the software will not prompt the user to shift the couch to the new isocenter. As a… | Terminated |
| Dec 24, 2015 | Z-0806-2016The system loaded a completion fraction in the incorrect order after a treatment interruption. | Terminated |
| Aug 27, 2015 | Z-0115-2016ViewRay discovered that in the event that an encoder breaks or fails on the Patient Handling System (PHS, or couch), when attempting to restart the system, the couch could move unexpectedly. | Terminated |
| May 7, 2015 | Z-1954-2015The software was failing to determine new patient locations if imaging is not enabled during treatment. | Terminated |
| Apr 1, 2015 | Z-2085-2015ViewRay received a report that the couch moved unexpectedly into the bore after performing a RTCS reboot. | Terminated |
| Jan 15, 2015 | Z-1580-2015The software was not correctly using the RT (Radiation Therapy) to MR (Magnetic Resonance image) coordinate correction for non HFS (Head First Supine) patient orientations, resulting in slice mismatch error. | Terminated |
- AerFrame™ Alignment Laser Box (LBV2)K261865 · Stabilix, LLC · Sep 16, 2026
- EZFluence (RADEZ V2)K261718 · Radformation, Inc. · Aug 31, 2026
- MRIA Radiation Therapy Information System (MRIA)K260733 · Manteia Technologies Co., Ltd. · Aug 26, 2026
- Halcyon, Ethos Radiotherapy System (5.0)K261856 · Varian Medical Systems, Inc. · Aug 3, 2026
- ThinkQA 3K252937 · Dosisoft SA · Jun 12, 2026
- Luna 3DK253004 · Lap GmbH Laser Applikationen · Jun 4, 2026
- AerFrame Patient Immobilization System (AFPIS)K260792 · Stabilix, LLC · May 8, 2026
- Veriqa RT Epid 3DK252258 · Ptw-Freiburg Physikalisch-Technische-Werkstaetten Dr. Pychla · Apr 15, 2026
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Jun 30, 2011
- Review panel
- Radiology
- Regulation
- 21 CFR 892.5050
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files