Mobilediagnost Wdr
System, X-Ray, Mobile. 510(k) K111725, from Sedecal., Sa. (Seneca, SC). Product code IZL, class 2.
As of Nov 18, 2026, K111725 is an FDA 510(k) held by Sedecal., Sa., substantially equivalent on Jul 19, 2011 after 29 days, with 8 recalls tied to it in FDA's recall database.
- Substantially equivalent: Jul 19, 2011
- Review time: 29 days
- Recalls: 8
- Cited as a predicate by: 1 device
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFIntended for use by a qualified/trained doctor or technologist on both adult and pediatric patients for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with patient sitting, standing or lying in the prone or supine positions. Not intended for mammography.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Easy Moving Plus, Sm-Xxhf-YyK090322 · Sedecal., Sa. · Mar 17, 2009
- Wireless Portable Detector FD-W17K090625 · Philips Medical Systems North America Co. · Mar 24, 2009
Cited as a Predicate
1 later 510(k) summary names this device as a predicate- Mobilediagnost WdrK141895 · Sedecal., Sa. · Sep 18, 2014
Recalls
Recalls FDA ties to K111725, newest first; each opens FDA's recall record| Jun 10, 2025 | Z-0243-2026Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences… | Open, Classified |
| Oct 14, 2021 | Z-1083-2024While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning… | Open, Classified |
| Jan 18, 2016 | Z-1084-2016This is related to previous 806 Notifications about uncommanded movements of a motorized mobile x-ray system. Reason for Recall: Along with the previously noted encoder and gauges problems and corresponding fixes, we… | Terminated |
| Mar 23, 2015 | Z-1691-2016Due to a software defect, the system may sporadically apply the default x ray exposure parameters for an adult ( patient type : Normal ) even though the patient type " Newborn" was selected and is displayed in the… | Terminated |
| Aug 6, 2014 | Z-1040-2015Reports of unexpected movement. | Terminated |
| Apr 8, 2014 | Z-0508-2015In special cases, during the start-up of the current model of MobileDiagnost wDR(WmDR 1.1), an unintended exposure could occur. | Terminated |
| Jun 11, 2013 | Z-1964-2013The Instructions for Use (IFU) for the MobileDiagnost wDR fails to comply with a Federal standard. The IFU does not provide tolerance levels for several generator values as required by 1020.30(h)(3). | Terminated |
| Nov 28, 2012 | Z-1706-2017When the operator moves an image to another view then the electronic marker might be placed at the wrong anatomical position in the resulting image. | Terminated |
Other IZL Clearances
The newest 510(k)s for the same product code; their median review was 166 days; see every IZL device- REMEX-T(K)100CK261943 · Remedi, Inc. · Oct 2, 2026
- XliteK260236 · Gemss Healthcare Co., Ltd. · Aug 3, 2026
- Portable DR Imaging SystemK253058 · Shantou Institute of Ultrasonic Instruments Co., Ltd. · Jun 18, 2026
- RAYMO Mobile X-ray System (Model: RAYMO)K260085 · DRGEM Corporation · May 14, 2026
- Lux HD Detectors, DRX-LC Detector, DRX-Revolution Mobile X-Ray SystemK253185 · Carestream Health · Mar 11, 2026
- AirRay-mini 80A, AirRay-mini 90A, AirRay-mini 80F, AirRay-mini 90F (AirRay-mini)K253244 · Poskom Co., Ltd. · Feb 23, 2026
- Airbile-100K250976 · Poskom Co., Ltd. · Dec 19, 2025
- Mobile X-ray unit (!M1)K251710 · Solutions For Tomorrow AB · Oct 24, 2025
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jun 20, 2011
- Review panel
- Radiology
- Regulation
- 21 CFR 892.1720
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files