Arca
Skip to content
Tools
Tools

Pinnacle Precision Access System

Introducer, Catheter. 510(k) K111606, from Terumo Medical Corp. (Elkton, MD). Product code DYB, class 2.

As of Nov 20, 2026, K111606 is an FDA 510(k) held by Terumo Medical Corp., substantially equivalent on Oct 3, 2011 after 116 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 116 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Pinnacle Precision Access System is used to facilitate placing a catheter through the skin into a vein or artery. The Entry Needle is an accessory device which is used to gain access to the vein or artery, for placement of the Mini Guide Wire. The Mini Guide Wire is an accessory device which is used for placement of the sheath into the vein or artery.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K111606, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Mar 16, 2016Z-2738-2016Terumo Medical is initiating this recall due to complaints of valve leakage that have been related to a manufacturing issue. If affected introducer sheaths are used, there is possible risk of incremental blood loss…Terminated

Other DYB clearances

The newest 510(k)s for the same product code; their median review was 71 days; see every DYB device

Record

Type
Traditional
Received
Jun 9, 2011
Review panel
Cardiovascular
Regulation
21 CFR 870.1340

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files