Nt 2000 Lesioning Generator

Generator, Lesion, Radiofrequency. 510(k) K111576, from Neuro Therm, Inc. (Ashland, MA). Product code GXD, class 2.

As of Nov 18, 2026, K111576 is an FDA 510(k) held by Neuro Therm, Inc., substantially equivalent on Sep 20, 2011 after 106 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Sep 20, 2011
  • Review time: 106 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K111576, newest first; each opens FDA's recall record
Sep 28, 2023Z-0209-2024Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.Open, Classified
Apr 22, 2021Z-1549-2021Potential capacitor failure which typically manifests as in inability to provide output on one or more channels. In some instances, users reported a charring/smoky odor or emission.Terminated
Oct 14, 2019Z-0664-2020Affected lots were manufactured with the protective release liner in an incorrect orientation. When attempting to remove this disposable release liner, a clear film may be left on the grounding pad which may interfere…Terminated
Oct 9, 2013Z-1361-2016NT2000iX software shipped with the international setting turned on. The international settings include access to Corodotomy, Bi-Polar, and No Temperature modes were not cleared in the United States for use.Terminated

Other GXD Clearances

The newest 510(k)s for the same product code; their median review was 268 days; see every GXD device

Record

Type
Traditional
Received
Jun 6, 2011
Review panel
Neurology
Regulation
21 CFR 882.4400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.