Arca
Skip to content
Tools
Tools

Ipc Powerease System

Drills, Burrs, Trephines & Accessories (Simple, Powered). 510(k) K111520, from Medtronic Xomed, Inc. (Jacksonville, FL). Product code HBE, class 2.

As of Nov 18, 2026, K111520 is an FDA 510(k) held by Medtronic Xomed, Inc., substantially equivalent on Oct 26, 2011 after 146 days, with 2 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 26, 2011
  • Review time: 146 days
  • Recalls: 2
  • Cited as a predicate by: 12 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The IPC ® System is indicated for the incision / cutting, removal, drilling, and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Cranial, Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

12 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K111520, newest first; each opens FDA's recall record
Jul 17, 2025Z-2461-2025Due to out of the box wobble of the driver.Open, Classified
Dec 3, 2014Z-0952-2015The retaining tabs component of the Set Screw Retaining Drive may be oversized, which will not allow proper mating between the Set Screw Retaining Drive and the set screw.Terminated

Other HBE Clearances

The newest 510(k)s for the same product code; their median review was 89 days; see every HBE device

Record

Type
Traditional
Received
Jun 2, 2011
Review panel
Neurology
Regulation
21 CFR 882.4310

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.