Arca
Skip to content
Tools
Tools

Liat (Tm) Influenza A / B Assay, Liat (Tm) Analyzer

Respiratory Virus Panel Nucleic Acid Assay System. 510(k) K111387, from Iquum, Inc. (Marlborough, MA). Product code OCC, class 2.

As of Nov 20, 2026, K111387 is an FDA 510(k) held by Iquum, Inc., substantially equivalent on Aug 4, 2011 after 79 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 79 days
  • Recalls: 4
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The IQuum Liat™ Influenza A/B Assay performed on the Liat™ Analyzer is an automated multiplex real-time RT-PCR assay for the rapid in vitro qualitative detection and discrimination of influenza A virus and influenza B virus RNA in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K111387, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Nov 16, 2021Z-0828-2022Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases…Open
May 17, 2016Z-2203-2016Kit labeled with the incorrect expiration date.Terminated
May 9, 2016Z-2216-2016Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid…Terminated
Jul 24, 2015Z-2543-2015Due to a supplier issue, the incorrect green and amber LED components were used in the manufacture of the LED printed circuit board assemblies (PCBA) in certain cobas Liat Analyzers.Terminated

Other OCC clearances

The newest 510(k)s for the same product code; their median review was 29 days; see every OCC device

Record

Type
Traditional
Received
May 17, 2011
Review panel
Microbiology
Regulation
21 CFR 866.3980

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.