Liat (Tm) Influenza A / B Assay, Liat (Tm) Analyzer
Respiratory Virus Panel Nucleic Acid Assay System. 510(k) K111387, from Iquum, Inc. (Marlborough, MA). Product code OCC, class 2.
As of Nov 20, 2026, K111387 is an FDA 510(k) held by Iquum, Inc., substantially equivalent on Aug 4, 2011 after 79 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 79 days
- Recalls: 4
- Cited as a predicate by: 3 devices
The IQuum Liat™ Influenza A/B Assay performed on the Liat™ Analyzer is an automated multiplex real-time RT-PCR assay for the rapid in vitro qualitative detection and discrimination of influenza A virus and influenza B virus RNA in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors.
- Xpert Flu, Genexpert DX Systems, Genexpert Infinity System, Genexpert System Software Verisions 2.1 and 4.0K103766 · Cepheid · Apr 21, 2011
- Cobas Influenza A/B Nucleic Acid Test for Use on the Cobas Liat SystemK191729 · Roche Molecular Systems, Inc. · Jul 24, 2019
- cobas Influenza A/B & RSV Nucleic Acid Test for Use on the cobas Liat SystemK153544 · Iquum, Inc. · Jul 25, 2016
- Alere I Influenza A & BK141520 · Alere Scarborough, Inc D/B/A Binax, Inc. · Jun 13, 2014
| Nov 16, 2021 | Z-0828-2022Roche has identified a small number of cobas Liat analyzer units that have consistently high noise levels (background signal) originating from the amber detector (used for Influenza B detection). This noise increases… | Open |
| May 17, 2016 | Z-2203-2016Kit labeled with the incorrect expiration date. | Terminated |
| May 9, 2016 | Z-2216-2016Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid… | Terminated |
| Jul 24, 2015 | Z-2543-2015Due to a supplier issue, the incorrect green and amber LED components were used in the manufacture of the LED printed circuit board assemblies (PCBA) in certain cobas Liat Analyzers. | Terminated |
- SimplicityDx Panel FluA/FluB/RSVK262789 · Molecular Designs · Nov 4, 2026
- ID NOW Influenza A & B 2; ID NOW RSV; ID NOW Strep A 2; ID NOW COVID-19 2.0K262747 · Abbott Diagnostics Scarborough, Inc. · Aug 31, 2026
- BioCode® Respiratory Pathogen Panel (RPP)K254139 · Applied BioCode, Inc. · May 1, 2026
- NxTAG® Respiratory Pathogen PanelK242613 · Luminex Molecular Diagnostics, Inc. · Oct 2, 2024
- ID NOW Influenza A & B 2K232775 · Abbott Diagnostics Scarborough, Inc. · Oct 10, 2023
- Panther Fusion AdV/hMPV/RV AssayK231017 · Hologic, Inc. · May 5, 2023
- cobas Influenza A/B & RSV nucleic acid test for use on the cobas Liat SystemK213822 · Roche Molecular Systems, Inc. · Jul 6, 2022
- cobas Influenza A/B & RSV Nucleic acid test for use on the cobas Liat SystemK210234 · Roche Molecular Systems, Inc. · Feb 16, 2021
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- May 17, 2011
- Review panel
- Microbiology
- Regulation
- 21 CFR 866.3980
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files