Arca
Skip to content
Tools
Tools

Accu-Check Combo System

Pump, Infusion, Insulin. 510(k) K111353, from Roche Diagnostics (Indianapolis, IN). Product code LZG, class 2.

As of Nov 20, 2026, K111353 is an FDA 510(k) held by Roche Diagnostics, substantially equivalent on Jul 17, 2012 after 428 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 428 days
  • Recalls: 3
  • Cited as a predicate by: 3 devices

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K111353, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Apr 22, 2015Z-1733-2015Roche has become aware of an increase of mechanical errors experienced by customers using the Accu-Chek Spirit Combo insulin infusion pump and Accu-Chek Spirit insulin infusion pump when the insulin pumps display E6 and…Terminated
Sep 25, 2014Z-0146-2015The insulin pump may lose time and date settings during a power interruption (such as a battery change) due to a faulty capacitor. If the capacitor fails, the time and date will return to default setting. Though the…Terminated
Oct 25, 2013Z-0385-2014There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus…Terminated

Other LZG clearances

The newest 510(k)s for the same product code; their median review was 146 days; see every LZG device

Record

Type
Abbreviated
Received
May 16, 2011
Review panel
General Hospital
Regulation
21 CFR 880.5725

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files