Arca
Skip to content

Issys LP Spinal Fixation System

Thoracolumbosacral Pedicle Screw System. 510(k) K111099, from Custom Spine, Inc. (Parsippany, NJ). Product code NKB, class 2.

As of Nov 17, 2026, K111099 is an FDA 510(k) held by Custom Spine, Inc., substantially equivalent on May 5, 2011 after 15 days, with 1 recall tied to it in FDA's recall database.

  • Substantially equivalent: May 5, 2011
  • Review time: 15 days
  • Recalls: 1

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The ISSYS LP Spinal Fixation System is intended to help provide immobilization and stabilization of the spinal segments as an adjunct to fusion of the lumbar and/or sacral spine, specifically as follows: For pedicular use: When used as pedicle screw fixation system of the non cervical posterior spine in skeletally mature patients, these systems are indicated for one or more of the following…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K111099, newest first; each opens FDA's recall record
StartedReasonStatus
Mar 28, 2013Z-1819-2013On March 28, 2013, Custom Spine, Inc. issued a minor removal of a lot of 18 screws that were found to be out of specification. The screw length was 5 mm shorter in length than the required specification for the screw…Terminated

Other NKB Clearances

The newest 510(k)s for the same product code; their median review was 90 days; see every NKB device

Record

Type
Special
Received
Apr 20, 2011
Review panel
Orthopedic
Regulation
21 CFR 888.3070

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.