Arca
Skip to content
Tools
Tools

Centricity Pacs

System, Image Processing, Radiological. 510(k) K110875, from GE Healthcare (Barrington, IL). Product code LLZ, class 2.

As of Nov 20, 2026, K110875 is an FDA 510(k) held by GE Healthcare, substantially equivalent on May 5, 2011 after 36 days, with 10 recalls tied to it in FDA's recall database.

  • Review time: 36 days
  • Recalls: 10
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Centricity PACS provides scalable image and data management solutions for medical imaging modalities, such as Computed Tomography (CT), Magnetic Resonance (MR), Computed Radiography (CR), Digital X-Ray (DX), Digital Mammography (MG), Ultrasound (US), Nuclear Medicine (NM), Positron Emission Tomography (PET), X-Ray Angiography (XA), Oral X-Ray (IO), Endoscopic Video (ES), and any other DICOM…

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K110875, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jul 31, 2026Z-3171-2026, Z-3169-2026, Z-3166-2026, Z-3167-2026, Z-3170-2026, Z-3168-2026Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient…Open
Aug 5, 2024Z-3214-2024GE HealthCare has become aware of an issue where image storage is not possible when Centricity PACS reaches the maximum number of unique image identifiers (Ckeys). If this occurs, it can result in acquisition of a…Open
Jun 14, 2022Z-1466-2022The Event Notification Manager (ENM) functionality for certain products does not process notifications for study modifications performed on post verified exams.Terminated
Apr 28, 2022Z-1233-2022Potential for Distance and Area measurements to display inaccurate measurement values when performed on magnified images, and specific to ZFP, Distance and Area measurements can display inaccurate measurement values…Open
May 16, 2019Z-1911-2019There is the possibility of viewing studies directly from the Enterprise Archive or VNA with incorrect patient images because the updated series or study is not present in the archive.Terminated

Other LLZ clearances

The newest 510(k)s for the same product code; their median review was 145 days; see every LLZ device

Record

Type
Traditional, third-party review
Received
Mar 30, 2011
Review panel
Radiology
Regulation
21 CFR 892.2050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files