Arca
Skip to content
Tools
Tools

Onetouch Verio (Tm) Iq Blood Glucose Monitoring System

Glucose Dehydrogenase, Glucose. 510(k) K110637, from Cilag GmbH International (Zug, CH). Product code LFR, class 2.

As of Nov 20, 2026, K110637 is an FDA 510(k) held by Cilag GmbH International, substantially equivalent on Sep 7, 2011 after 187 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 187 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The OneTouch® Verio™ IQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm or palm. The OneTouch® Verio™ IQ Blood Glucose Monitoring System is intended to be used by a single patient and should not be used for testing multiple patients.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K110637, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Feb 21, 2012Z-1359-2012An error was discovered in the OneTouch IQ meter software so that the meter turns itself off when a user attempts to view results in the "Results Log" when the log has 256 or a multiple of 256 items to display.Terminated

Record

Type
Traditional
Received
Mar 4, 2011
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1345

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files