Arca
Skip to content
Tools
Tools

Sound Surgical Vaser 2.1 Lipo System

System, Suction, Lipoplasty For Removal. 510(k) K110306, from Sound Surgical Technologies, LLC (Louisville, CO). Product code QPB, class 2.

As of Nov 19, 2026, K110306 is an FDA 510(k) held by Sound Surgical Technologies, LLC, substantially equivalent on Jun 23, 2011 after 141 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 141 days
  • Recalls: 1
  • Cited as a predicate by: 3 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The Sound Surgical VASER 2.1 is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring. The Sound Surgical VentX 2.1 is intended for the suction or aspiration of fluids and tissue during surgical procedures. The VentX 2.1 is designed to operate with the Sound Surgical VASER 2.1 or as a stand-alone system.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

3 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K110306, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Jun 1, 2017Z-3047-2017Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.Terminated

Other QPB clearances

The newest 510(k)s for the same product code; their median review was 220 days; see every QPB device

Record

Type
Traditional
Received
Feb 2, 2011
Review panel
General, Plastic Surgery
Regulation
21 CFR 878.5040

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files