Nidek Multicolor Laser Photocoagulator
Photocoagulator And Accessories. 510(k) K110228, from Nidek Incorporated (Sunnyvale, CA). Product code HQB, class 2.
As of Nov 19, 2026, K110228 is an FDA 510(k) held by Nidek Incorporated, substantially equivalent on Feb 23, 2011 after 28 days, with 4 recalls tied to it in FDA's recall database.
- Review time: 28 days
- Recalls: 4
- Cited as a predicate by: 3 devices
The Nidek Multicolor Laser Photocoagulator System MC-500 is intended to be used in ophthalmic surgical procedures, including retinal and macular photocoagulation, iridotomy and trabeculoplasty.
- Multi Color Laser Photocoagulator, Model MC-300K042785 · Nidek, Inc. · Dec 14, 2004
- Green Laser Photocoagulator GYC-500K152603 · Nidek Co., Ltd. · May 2, 2016
- Nidek Multicolor Laser Photocoagulator System MC-500 Connected With The Endophotocoagulation Delivery UnitK133358 · Nidek Co., Ltd. · Aug 12, 2014
- Opto Mitra Yellow LaserK111643 · Opto Electronica SA · Nov 3, 2011
| Jun 9, 2014 | Z-2132-2014Laser aperture label was not affixed to device prior to shipment in the US. | Terminated |
| May 24, 2013 | Z-0222-2015Retrospective review found that a Engineering Change Order initiated in May 2013 for a memory function problem was not reported as required. | Terminated |
| Oct 24, 2012 | Z-0224-2015When the MC-500 is turned on and coagulation is performed without changing the laser color, the delivery unit occasionally fails to emit the laser, or the power output is lower than the set program. | Terminated |
| Mar 27, 2012 | Z-1127-2015The spot size control of the scan delivery unit of the MC-500 Multicolor Laser Photocoagulator may turn by itself. | Terminated |
- Vortex Surgical Laser Probes, Vortex Surgical Illuminated Laser Probes, Vortex Surgical Endoilluminators, Vortex…K220263 · Vortex Surgical, Inc. · Nov 3, 2022
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Jan 26, 2011
- Review panel
- Ophthalmic
- Regulation
- 21 CFR 886.4690
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files