Arca
Skip to content

Optima CT660

System, X-Ray, Tomography, Computed. 510(k) K110227, from Ge Medical Systems, LLC (Waukesha, WI). Product code JAK, class 2.

As of Nov 17, 2026, K110227 is an FDA 510(k) held by Ge Medical Systems, LLC, substantially equivalent on Jul 20, 2011 after 176 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jul 20, 2011
  • Review time: 176 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Optima CT660 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate
  • Optima CT660K131576 · Ge Medical Systems, LLC · Aug 30, 2013

Recalls

Recalls FDA ties to K110227, newest first; each opens FDA's recall record
May 24, 2018Z-2583-2018Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).Terminated
Apr 27, 2018Z-2873-2018GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.Terminated
Apr 27, 2018Z-2871-2018GE Healthcare has identified that some CT systems may have a damaged cable that can expose 120VAC to Service Engineers working inside the gantry.Terminated
May 19, 2017Z-2182-2017It was discovered that an issue has been identified that can result in a scan starting before the prescribed delay has completed.Terminated
Mar 26, 2013Z-0023-2014GE Healthcare has recently become aware of a potential safety issue related with the DMPR function and Screen Saved Images on your scanner which involves Optima CT660, Brivo CT385, Discovery PET/CT 610, Discovery PET/CT…Terminated

Other JAK Clearances

The newest 510(k)s for the same product code; their median review was 237 days; see every JAK device

Record

Type
Special
Received
Jan 25, 2011
Review panel
Radiology
Regulation
21 CFR 892.1750

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.