Tester, Defibrillator. 510(k) K110192, from Bc Group International, Inc. (Oak Grove, MN). Product code DRL, class 2.

As of Nov 19, 2026, K110192 is an FDA 510(k) held by Bc Group International, Inc., substantially equivalent on Feb 8, 2011 after 15 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 15 days
  • Recalls: 1
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The DA-2006/DA-2006P are used to determine that defibrillators and transcutaneous pacemakers are performing within their performance specifications through the measurement of energy output. The DA-2006 tests defibrillators while the DA-2006P additionally tests transcutaneous pacemakers.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate
  • Delta 3300K190437 · Netech Corporation · Aug 28, 2019

Recalls

Recalls FDA ties to K110192, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Oct 28, 2015Z-0457-2016Device is not functioning as intended: Two wires running to the ECG distribution circuit board are reversed resulting in inverted waveforms.Terminated

Record

Type
Traditional, third-party review
Received
Jan 24, 2011
Review panel
Cardiovascular
Regulation
21 CFR 870.5325

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files