Empira RX Ptca Dilatation Catheter, Empria NC RX Ptca Dilatation Catheter

Catheters, Transluminal Coronary Angioplasty, Percutaneous. 510(k) K110133, from Creganna-Tactx Medical (Campbell, CA). Product code LOX, class 2.

As of Nov 20, 2026, K110133 is an FDA 510(k) held by Creganna-Tactx Medical, substantially equivalent on Jun 10, 2011 after 143 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 143 days
  • Recalls: 3
  • Cited as a predicate by: 5 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

"The EMPIRA Rx PTCA Catheter and EMPIRA NC Rx PTCA Dilatation Catheter are indicated for balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis for the purpose of improving myocardial perfusion. The EMPIRA NC Rx PTCA Dilatation Catheter is also indicated for post-delivery expansion of balloon expandable stents."

Cited as a predicate

5 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K110133, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 23, 2014Z-2557-2014, Z-2556-2014The balloon may exhibit resistance to the removal of the balloon cover. Weakening of the affected product in the region of the proximal balloon seal may occur when excessive force is applied to remove the balloon cover.…Terminated
Feb 5, 2013Z-1784-2014Dilatation catheters could exhibit radial versus axial tears should they burst during inflation.Terminated

Other LOX clearances

The newest 510(k)s for the same product code; their median review was 205 days; see every LOX device

Record

Type
Traditional
Received
Jan 18, 2011
Review panel
Cardiovascular
Regulation
21 CFR 870.5100

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files