Ventilator, Continuous, Facility Use. 510(k) K103625, from Draeger Medical Systems, Inc. (Telford, PA). Product code CBK, class 2.

As of Nov 17, 2026, K103625 is an FDA 510(k) held by Draeger Medical Systems, Inc., substantially equivalent on Oct 6, 2011 after 300 days, with 4 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Oct 6, 2011
  • Review time: 300 days
  • Recalls: 4
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Oxylog 3000 plus is a time-cycled, volume controlled and pressure controlled emergency and transport ventilator for patients requiring mandatory or assisted ventilation with a tidal volume from 50 mL upwards. *August 2011* *Page 2 of 4* *-* *Oxylog 3000 plus* *K103625* *Dräger*

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K103625, newest first; each opens FDA's recall record
Jun 12, 2023Z-2018-2023Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.Open, Classified
Dec 22, 2015Z-0692-2016Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.Terminated
Dec 22, 2015Z-0691-2016Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.Terminated
Mar 5, 2014Z-1440-2014The potentiometers (knobs) for setting the ventilation parameters may have contact interrupted and the device discontinues ventilation, gives an audible alarm and displays the error message Poti unplugged.Terminated

Other CBK Clearances

The newest 510(k)s for the same product code; their median review was 235 days; see every CBK device

Record

Type
Traditional
Received
Dec 10, 2010
Review panel
Anesthesiology
Regulation
21 CFR 868.5895

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files