Arca
Skip to content

Dx-D 100

System, X-Ray, Mobile. 510(k) K103597, from Agfa Healthcare N.V. (Greenville, SC). Product code IZL, class 2.

As of Nov 17, 2026, K103597 is an FDA 510(k) held by Agfa Healthcare N.V., substantially equivalent on Jan 3, 2011 after 26 days, with 6 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 3, 2011
  • Review time: 26 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

Agfa's DX-D 100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis. Systems can be used with MUSICA 2 image processing to create radiographic images of the skeleton (including skull, spinal column and extremities) chest, abdomen and other body parts. Agfa's DX-D 100 is not indicated for use in mammography.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

1 later 510(k) summary names this device as a predicate
  • DR 100sK191884 · Agfa N.V. · Aug 9, 2019

Recalls

Recalls FDA ties to K103597, newest first; each opens FDA's recall record
Feb 23, 2024Z-1462-2024Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is…Open, Classified
Jan 20, 2016Z-1698-2017The DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored.Terminated
May 29, 2015Z-1745-2015When using the DX-D 100 unit, sporadic unintended movements caused by an electrostatic discharge of the unit to the ground may occur.Terminated
Jun 6, 2014Z-2081-2014When liquid comes in contact with the DX-D 100 touch screen, the device may incorrectly recognize this as user input altering device settings.Terminated
May 13, 2014Z-2176-2014Unit was mislabeled with a factory label showing 32 KW output power.Terminated
May 15, 2013Z-1487-2013Customers could potentially experience intermittent, unintended and illogical movement when using the product.Terminated

Other IZL Clearances

The newest 510(k)s for the same product code; their median review was 166 days; see every IZL device

Record

Type
Traditional, third-party review
Received
Dec 8, 2010
Review panel
Radiology
Regulation
21 CFR 892.1720

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files