Arca
Skip to content
Tools
Tools

Neoblue Blanket Led Phototherapy System

Unit, Neonatal Phototherapy. 510(k) K103589, from Natus Medical Incorporated (Seattle, WA). Product code LBI, class 2.

As of Nov 18, 2026, K103589 is an FDA 510(k) held by Natus Medical Incorporated, substantially equivalent on May 13, 2011 after 157 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 157 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The neoBLUE blanket LED Phototherapy System is intended for the treatment of neonatal hyperbilirubinemia. It can be used in the clinical setting or in the home.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K103589, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 28, 2016Z-0770-2017neoBLUE blanket LED Phototherapy System distributed since release of the product in 2011 is recalled due to customer complaints regarding discoloration/ degradation of fiberoptic bundle at connection of pad and box.Terminated
Sep 16, 2016Z-0769-2017The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of…Terminated
Mar 16, 2015Z-1412-2015neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and…Terminated

Other LBI clearances

The newest 510(k)s for the same product code; their median review was 211 days; see every LBI device

Record

Type
Traditional
Received
Dec 7, 2010
Review panel
General Hospital
Regulation
21 CFR 880.5700

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.