Neoblue Blanket Led Phototherapy System
Unit, Neonatal Phototherapy. 510(k) K103589, from Natus Medical Incorporated (Seattle, WA). Product code LBI, class 2.
As of Nov 18, 2026, K103589 is an FDA 510(k) held by Natus Medical Incorporated, substantially equivalent on May 13, 2011 after 157 days, with 3 recalls tied to it in FDA's recall database.
- Review time: 157 days
- Recalls: 3
- Cited as a predicate by: 2 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe neoBLUE blanket LED Phototherapy System is intended for the treatment of neonatal hyperbilirubinemia. It can be used in the clinical setting or in the home.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Olympic Bili-Lite-Pad(Tm)K901987 · Olympic Medical Corp. · Jul 16, 1990
- Neoblue Cozy Led Phototherapy SystemK051869 · Natus Medical, Inc. · Oct 7, 2005
- Bilisoft Phototherapy SystemK053568 · Lumitex, Inc. · Apr 17, 2006
Cited as a predicate
2 later 510(k) summaries name this device as a predicate- neoBLUE blanket LED Phototherapy SystemK182178 · Natus Medical Incorporated · Dec 13, 2018
- Bilicocoon™ Phototherapy SystemK163526 · Neomedlight · Oct 6, 2017
Recalls
Recalls FDA ties to K103589, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Oct 28, 2016 | Z-0770-2017neoBLUE blanket LED Phototherapy System distributed since release of the product in 2011 is recalled due to customer complaints regarding discoloration/ degradation of fiberoptic bundle at connection of pad and box. | Terminated |
| Sep 16, 2016 | Z-0769-2017The firm sent the Technical Bulletin to consignees who received the reliability updated neoBLUE blanket LED Phototherapy Systems between March 2016 to August 2016. The reliability update to prevent the early failure of… | Terminated |
| Mar 16, 2015 | Z-1412-2015neoBLUE blanket LED Phototherapy System is recalled because Natus has received feedback from some customers about early failure of the neoBLUE blanket system fiberoptic pads. These failures involve discoloration and… | Terminated |
Other LBI clearances
The newest 510(k)s for the same product code; their median review was 211 days; see every LBI device- SnugLit(TM) Wearable Phototherapy System (SNGL-01-US)K251308 · Thera B Medical Products · Jan 15, 2026
- BiliWrapK243372 · Gerium Medical, Ltd. · May 29, 2025
- Infant Phototherapy EquipmentK210289 · Bistos Co., Ltd. · May 28, 2021
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Dec 7, 2010
- Review panel
- General Hospital
- Regulation
- 21 CFR 880.5700
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files