Arca
Skip to content

Skyline Anterior Cervical Plate System

Appliance, Fixation, Spinal Intervertebral Body. 510(k) K103491, from A Johnson & Johnson Company (Raynham, MA). Product code KWQ, class 2.

As of Nov 17, 2026, K103491 is an FDA 510(k) held by A Johnson & Johnson Company, substantially equivalent on Feb 14, 2011 after 80 days, with no recalls tied to it in FDA's recall database.

  • Substantially equivalent: Feb 14, 2011
  • Review time: 80 days
  • Recalls: 0
  • Cited as a predicate by: 12 devices

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The SKYLINE® Anterior Cervical Plate System is indicated for stabilization of the cervical spine from C2 to C7 employing unicortical screw fixation at the anterior face of the vertebral bodies. Specific clinical indications for anterior plating include: Instability caused by trauma; instability associated with correction of cervical lordosis and kyphosis deformity; instability associated with…

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a Predicate

12 later 510(k) summaries name this device as a predicate

Other KWQ Clearances

The newest 510(k)s for the same product code; their median review was 86 days; see every KWQ device

Record

Type
Special
Received
Nov 26, 2010
Review panel
Orthopedic
Regulation
21 CFR 888.3060

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.