Spacelabs Multi-Parameter Module
Detector And Alarm, Arrhythmia. 510(k) K103142, from Spacelabs Healthcare, Inc. (Nederland, CO). Product code DSI, class 2.
As of Nov 18, 2026, K103142 is an FDA 510(k) held by Spacelabs Healthcare, Inc., substantially equivalent on Mar 3, 2011 after 129 days, with 14 recalls tied to it in FDA's recall database.
- Review time: 129 days
- Recalls: 14
- Cited as a predicate by: 1 device
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Spacelabs Medical Multiparameter Module 91496K050605 · Spacelabs Medical, Inc. · May 3, 2005
Cited as a predicate
1 later 510(k) summary names this device as a predicate- ePM Series Patient monitorsK200015 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Jun 8, 2020
Recalls
Recalls FDA ties to K103142, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Apr 24, 2026 | Z-2247-2026, Z-2246-2026, Z-2243-2026, Z-2245-2026, Z-2244-2026GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF… | Open |
| Mar 26, 2026 | Z-2052-2026Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester). | Open |
| Feb 27, 2026 | Z-1793-2026Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO… | Open |
| Apr 12, 2023 | Z-1503-2023Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent… | Open |
| Jul 1, 2016 | Z-2343-2016The firm has received multiple reports of the Non-Invasive Blood Pressure (NIBP) parameter becoming non-functional with an associated message (No Reading). In addition, there is one of the following three error… | Terminated |
| Apr 21, 2015 | Z-1577-2015Modules resetting. This failure mode happens when the Spacelabs Command Module is configured with the Masimo SpO2 option (-M), the Spacelabs Respiration option (-R) and the Masimo SpO2 PCBA, PN: 010-1136-02. With these… | Terminated |
| Apr 17, 2015 | Z-1542-2015Ultraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M was affected. Customer reported that during the time that the module is resetting (~10 seconds), monitoring of all parameters… | Terminated |
| May 8, 2014 | Z-1705-2014A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms. | Terminated |
| Nov 3, 2011 | Z-0690-2012NIBP pump inside the Ultraview SL Command Module, Model 91496, became loose. The pump is suspended by the air hoses and electrical connections, but there can be enough movement to damage internal components and cause… | Terminated |
| Jul 13, 2011 | Z-0776-2012One Spacelabs Medical Ultraview SL Command Module, Model 91496, was removed from customer due to software error. | Terminated |
Other DSI clearances
The newest 510(k)s for the same product code; their median review was 248 days; see every DSI device- BodyGuardian Remote Monitoring System (BGRMS v3.0)K243349 · Boston Scientific Cardiac Diagnostic Technologies, Inc. · Jan 23, 2026
- MoMe ARC® Wireless Ambulatory ECG Monitoring and Detection System (32000)K250356 · Infobionic, Inc. · Jul 29, 2025
- Zio AT® device (A100A1001)K240177 · iRhythm Technologies, Inc. · Oct 30, 2024
- MoMe® ARC Wireless Ambulatory ECG Monitoring and Detection SystemK230265 · Infobionic, Inc. · Oct 6, 2023
- SmartCardia 7L PlatformK231276 · Smartcardia SA · Aug 30, 2023
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Oct 25, 2010
- Review panel
- Cardiovascular
- Regulation
- 21 CFR 870.1025
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files