Arca
Skip to content
Tools
Tools

Forza Spacer System

Intervertebral Fusion Device With Bone Graft, Lumbar. 510(k) K103111, from Orthofix, Inc. (Lewisville, TX). Product code MAX, class 2.

As of Nov 18, 2026, K103111 is an FDA 510(k) held by Orthofix, Inc., substantially equivalent on Mar 23, 2011 after 153 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 153 days
  • Recalls: 1
  • Cited as a predicate by: 7 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The FORZA™ Spacer System is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels in the lumbar spine (L2-S1). DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

7 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K103111, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Feb 6, 2012Z-1037-2012The 8mm and 10mm Rotating Cutters may fracture during use, specifically during disc space preparation. If an 8mm or 10mm Rotating Cutter does Fracture during use, it may result in a surgical delay estimated to be 30…Terminated

Other MAX clearances

The newest 510(k)s for the same product code; their median review was 142 days; see every MAX device

Record

Type
Traditional
Received
Oct 21, 2010
Review panel
Orthopedic
Regulation
21 CFR 888.3080

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files