Miradry System
Instrument For Treatment Of Hyperhidrosis. 510(k) K103014, from Miramar Labs, Inc. (Sunnyvale, CA). Product code OUB, class 2.
As of Nov 20, 2026, K103014 is an FDA 510(k) held by Miramar Labs, Inc., substantially equivalent on Jan 28, 2011 after 107 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 107 days
- Recalls: 2
- Cited as a predicate by: 2 devices
The miraDry System is indicated for use in the treatment of primary axillary hyperhidrosis. Note: The miraDry System is not indicated for treating hyperhidrosis related to other body areas or generalized hyperhidrosis. Predicate Device K082819 Miramar Labs' DTS G2 System
- Dermatologic Treatment System (Dts) G2K082819 · Miramar Labs, Inc. · Mar 20, 2009
- Sidelaze800 Laser Beam Delivery Accessory for Cynosure 1440MM Wavelength LasersK141511 · Cynosure, Inc. · Jan 23, 2015
- Miradry SystemK131162 · Miramar Labs, Inc. · Oct 25, 2013
| Oct 9, 2015 | Z-0234-2016Update to labeling for the miraDry Console to remind users to only use approved lubricants in a thin layer to prevent potential burns and blisters. | Terminated |
| Apr 24, 2011 | Z-2591-2011Reports of skin burns on axilla after treatment with the miraDry System. | Terminated |
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Oct 13, 2010
- Review panel
- General, Plastic Surgery
- Regulation
- 21 CFR 878.4400
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files