Arca
Skip to content
Tools
Tools

Intellisphere Event Managment System Rel. 10.00

System, Network And Communication, Physiological Monitors. 510(k) K102974, from Philips Medical Systems (Boca Raton, FL). Product code MSX, class 2.

As of Nov 20, 2026, K102974 is an FDA 510(k) held by Philips Medical Systems, substantially equivalent on Jan 4, 2011 after 90 days, with no recalls tied to it in FDA's recall database.

  • Review time: 90 days
  • Recalls: 0
  • Cited as a predicate by: 4 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The intended use of the Intellisphere Event Management product is to provide healthcare professionals with supplemental information about patient alarms, technical alarms, nurse call alarms and system information messages (events). The product can route all or subsets of this information to selective remote devices such as pagers, phones, or marquees.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

4 later 510(k) summaries name this device as a predicate

Other MSX clearances

The newest 510(k)s for the same product code; their median review was 266 days; see every MSX device

Record

Type
Traditional
Received
Oct 6, 2010
Review panel
Cardiovascular
Regulation
21 CFR 870.2300

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.