Arca
Skip to content
Tools
Tools

Torflex Transseptal Guiding Sheath

Introducer, Catheter. 510(k) K102948, from Baylis Medical Co., Inc. (Mississauga, Ontario, CA). Product code DYB, class 2.

As of Nov 19, 2026, K102948 is an FDA 510(k) held by Baylis Medical Co., Inc., substantially equivalent on Apr 1, 2011 after 178 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 178 days
  • Recalls: 3
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The TorFlex™ Transseptal Guiding Sheath is used for the percutaneous introduction of various types of cardiovascular catheters to all heart chambers, including the left atrium via transseptal perforation / puncture.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K102948, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Oct 7, 2019Z-2070-2020There is a remote potential for the sterile packaging of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified and the product is used in a…Terminated
Oct 7, 2019Z-2069-2020There is a remote potential for the sterile packaging of lots of the transseptal guiding sheath to be compromised during shipping and handling. In the case that a packaging defect is not identified and the product is…Terminated
Sep 23, 2013Z-0085-2014The TorFlex Transseptal Guiding Sheath Kits are being recalled due to the possibility that the sheath could remove particulate of less than 5 mm from the dilator. The particulate could potentially migrate into a…Terminated

Other DYB clearances

The newest 510(k)s for the same product code; their median review was 71 days; see every DYB device

Record

Type
Traditional
Received
Oct 5, 2010
Review panel
Cardiovascular
Regulation
21 CFR 870.1340

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files