Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer. 510(k) K102795, from Materialise NV (Leuven, BE). Product code JWH, class 2.
As of Nov 17, 2026, K102795 is an FDA 510(k) held by Materialise NV, substantially equivalent on Feb 2, 2011 after 128 days, with 4 recalls tied to it in FDA's recall database.
- Substantially equivalent: Feb 2, 2011
- Review time: 128 days
- Recalls: 4
- Cited as a predicate by: 6 devices
Indications for Use
As the 510(k) summary states them, read from FDA's summary PDFSignature™ Personalized Patient Care System is intended to be used as a surgical instrument to assist in the positioning of total knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
Predicate Devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- SurgicaseK073449 · Materialise NV · Apr 16, 2008
- Zimmer Patient Specific Instruments System, Version 2.0K093533 · Materialise NV · Feb 17, 2010
Cited as a Predicate
6 later 510(k) summaries name this device as a predicate- Conformis BeneFIT Hip SystemK190904 · Conformis, Inc. · Aug 28, 2019
- Materialise TKA Guide System, Materialise TKA Planner, Materialise TKA GuidesK150928 · Materialise NV · Aug 10, 2015
- Signature Planner, Signature GuidesK140257 · Materialise NV · Apr 2, 2014
- Zimmer Patient Specific Instrument System Zimmer Patient Specific Instruments Zimmer…K113829 · Materialise NV · Apr 2, 2012
- Signature Planner,Signature GuidesK111863 · Materialise NV · Jun 15, 2012
- Signature Planner / Signature GuidesK110415 · Materialise NV · May 16, 2011
Recalls
Recalls FDA ties to K102795, newest first; each opens FDA's recall record| May 3, 2016 | Z-1941-2016A patient specific case of Signature Personalized Patient Care System Knee Guides included a femur guide which does not match the patient associated with the case. If the issue is not identified prior to surgery, the… | Terminated |
| Sep 23, 2015 | Z-0162-2016Potential inaccuracies in the alignment of the guides. Surgery was inaccurately planned because the patient image data was incorrectly processed. | Terminated |
| Feb 4, 2015 | Z-1287-2015Images belonging to a different patient were used for the production of the patient specific surgical guide. The incorrect guide may cause a delay in surgery to accommodate the preparation and use of traditional… | Terminated |
| Aug 4, 2011 | Z-1145-2012Biomet has initiated this action following investigation that identified two (2) specific lots of part number 42-422561 were incorrectly labeled. One (1) unit (lot 025640) was received back from distributor's office as… | Terminated |
Other JWH Clearances
The newest 510(k)s for the same product code; their median review was 88 days; see every JWH device- MicroPort Total Knee SystemK263065 · Suzhou Microport Orthorecon Co., Ltd. · Oct 29, 2026
- GMK ExtensionK262057 · Medacta International SA · Sep 15, 2026
- GSPE SpacerK260063 · The General Hospital Corporation, d/b/a Massachusetts · Sep 11, 2026
- EVOLUTION MPX Total Knee SystemK262293 · Suzhou Microport Orthorecon Co., Ltd. · Sep 10, 2026
- b-ONE® Mobio Total Knee System Revision Cemented Tibial Augments, Stem Extensions and…K262341 · B-One Ortho, Corp. · Sep 8, 2026
- ATTUNE Revision Knee SystemK262130 · Depuy Orthopaedics, Inc. · Sep 4, 2026
- SK Knee SystemK253961 · Just Medical Devices (Tianjin) Co., Ltd. · Sep 1, 2026
- Freedom Total Knee System (E-Poly Tibial Liners)K261563 · Maxx Orthopedics, Inc. · Aug 7, 2026
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Traditional
- Received
- Sep 27, 2010
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3560
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files