Multi-Analyte Controls, All Kinds (Assayed). 510(k) K102700, from Alere, San Diego, Dba Biosite INcorporated, Dba IN (San Diego, CA). Product code JJY, class 1.

As of Nov 28, 2026, K102700 is an FDA 510(k) held by Alere, San Diego, Dba Biosite INcorporated, Dba IN, substantially equivalent on Dec 21, 2010 after 92 days, with 3 recalls tied to it in FDA's recall database.

  • Review time: 92 days
  • Recalls: 3

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

This material is designed for use with any Alere Cholestech LDX® cassette type that includes total cholesterol, HDL cholesterol, triglycerides and glucose. 5.2. Summary of Changes from Predicate Device: The predicate devices are the Lipid Controls, Multianalyte Controls and Calibration Verification Materials manufactured by Streck for Alere.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF
  • Chem-ChexK913687 · Streck Laboratories, Inc. · Sep 19, 1991

Recalls

Recalls FDA ties to K102700, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Aug 27, 2015Z-0122-2016Alere San Diego is recalling the Alere Cholestech LDX Multianalyte Control because the assigned control ranges for the Total Cholesterol (TC) and Triglyceride (TRG) analytes are incorrect.Terminated
Oct 22, 2012Z-0747-2013, Z-0746-2013Out of range (increased) HDL Cholesterol results.Terminated

Record

Type
Traditional
Received
Sep 20, 2010
Review panel
Clinical Chemistry
Regulation
21 CFR 862.1660

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files