Depuy Pinnacle Altrx Acetabular Liners
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented. 510(k) K102423, from DePuy Orthopaedics, Inc. (Warsaw, IN). Product code LPH, class 2.
As of Nov 18, 2026, K102423 is an FDA 510(k) held by DePuy Orthopaedics, Inc., substantially equivalent on Oct 29, 2010 after 65 days, with 2 recalls tied to it in FDA's recall database.
- Review time: 65 days
- Recalls: 2
- Cited as a predicate by: 3 devices
Indications for use
As the 510(k) summary states them, read from FDA's summary PDFThe DePuy Pinnacle AltrX Acetabular Cup Liners are indicated for use in total hip replacement procedures. Total hip replacement is indicated in the following conditions: 1. A severely painful and/or disabled joint from osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia. 2. Avascular necrosis of the femoral head. 3.
Predicate devices
The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF- Depuy Pinnacle Altrx Acetabular LinersK072963 · DePuy Orthopaedics, Inc. · Jan 9, 2008
Cited as a predicate
3 later 510(k) summaries name this device as a predicate- DePuy SUMMIT Porocoat Hip Prosthesis – MR Conditional, DePuy SUMMIT DuoFix Hip Prosthesis – MR Conditional, DePuy…K231873 · Depuy Ireland UC · Jan 19, 2024
- MPACT ExtensionK183582 · Medacta International S.A. · Mar 21, 2019
- Depuy Pinnacle Altrx Acetabular LinersK132959 · DePuy Orthopaedics, Inc. · Mar 10, 2014
Recalls
Recalls FDA ties to K102423, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record| Jul 24, 2012 | Z-2220-2012In the recalled lot, the products were not machined according to the production router, a size 52 cup (122136052) was machined instead. This resulted in a size 52 liner being etched and labeled as a size 54 liner. | Terminated |
| Mar 7, 2012 | Z-1366-2013DePuy Orthopaedics, Inc. (DePuy) is issuing a voluntary recall of one lot of the AltrX" liner because the product was incorrectly labeled on the box. The box was labeled for a 36IDX52OD +4 Neutral Liner, but the product… | Terminated |
Other LPH clearances
The newest 510(k)s for the same product code; their median review was 282 days; see every LPH device- G7® Acetabular SystemK261453 · Zimmer Biomet · Sep 23, 2026
- Mpact 3D Metal FoamK253630 · Medacta International S.A. · Aug 28, 2026
- TRIVANCE™ Hip StemK254043 · Globus Medical, Inc. · Aug 4, 2026
- Omnifit® 20° Series I Insert;Omnifit® Crossfire® 10° Series II Insert; Constrained Acetabular Insert; Trident®…K261624 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · Jul 10, 2026
- iNSitu Lateralized Acetabular Liners, iNSitu Femoral Stems with Plasma Sprayed CPTi CoatingK243192 · Nextstep Arthopedix, LLC · Jun 5, 2026
- Taperloc® Complete Hip StemK251270 · Biomet, Inc. · May 19, 2026
- G7® Revision Acetabular SystemK252623 · Zimmer, Inc. · Feb 4, 2026
- Mpact 3D Metal Implants Extension – DMLS TechnologyK251043 · Medacta International S.A. · Jan 28, 2026
Elsewhere
- Japan approved Pinnacle Ceramic Liner (CERAMAX), the same brandPMDA · Mar 29, 2013
Record
- Document
- 510(k) summary (PDF) ↗
- Type
- Special
- Received
- Aug 25, 2010
- Review panel
- Orthopedic
- Regulation
- 21 CFR 888.3358
One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files