Incubator, Neonatal. 510(k) K102226, from Atom Medical Corporation (Bunkyo-Ku, Tokyo, JP). Product code FMZ, class 2.

As of Nov 18, 2026, K102226 is an FDA 510(k) held by Atom Medical Corporation, substantially equivalent on Jan 4, 2011 after 148 days, with 5 recalls tied to it in FDA's recall database.

  • Substantially equivalent: Jan 4, 2011
  • Review time: 148 days
  • Recalls: 5

Indications for Use

As the 510(k) summary states them, read from FDA's summary PDF

The Dual Incu i is a combination of an infant incubator and an infant warmer. The device can transition from one mode to the other on user's demand. The Dual Incu i is an incubator for newborns and premature neonates.

Predicate Devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Recalls

Recalls FDA ties to K102226, newest first; each opens FDA's recall record
May 1, 2012Z-1945-2012There is a potential for the device's canopy to move unintentionally when using the incubator mode or radiant warmer mode.Terminated
Aug 23, 2011Z-0337-2012The caster with lock on the incubator or infant warmer may become loose or break.Terminated
Aug 23, 2011Z-0338-2012The caster with lock on the incubator may become loose or break.Terminated
Aug 23, 2011Z-0339-2012The caster with lock on the incubator may become loose or break.Terminated
Aug 23, 2011Z-0340-2012The caster with lock on the radiant-warming open type incubator may become loose or break.Terminated

Other FMZ Clearances

The newest 510(k)s for the same product code; their median review was 179 days; see every FMZ device

Record

Type
Traditional
Received
Aug 9, 2010
Review panel
General Hospital
Regulation
21 CFR 880.5400

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Logos are each company's own site icon, matched by name through Wikidata.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.