Abutment, Implant, Dental, Endosseous. 510(k) K102209, from Biomet 3i, Inc. (Palm Beach Gardens, FL). Product code NHA, class 2.

As of Nov 19, 2026, K102209 is an FDA 510(k) held by Biomet 3i, Inc., substantially equivalent on Dec 3, 2010 after 120 days, with 5 recalls tied to it in FDA's recall database.

  • Review time: 120 days
  • Recalls: 5
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

BIOMET 3i Patient Specific Dental Abutments (Encode®) Designed Using Cadent iTero Scanner and inLab Software v3.5 are intended for use as an accessory to endosseous dental implants to support a prosthetic device in a partially or edentulous patient. These are intended for use to support single and multiple tooth prostheses, in the mandible or maxilla.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K102209, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Feb 26, 2019Z-2142-2019, Z-2140-2019, Z-2144-2019, Z-2143-2019, Z-2141-2019During manufacturing, the screw stop ledge was not made to specification and location within the abutment, resulting in the screw head seating deeper in the abutment with no positive stop.Terminated

Other NHA clearances

The newest 510(k)s for the same product code; their median review was 169 days; see every NHA device

Record

Type
Traditional
Received
Aug 5, 2010
Review panel
Dental
Regulation
21 CFR 872.3630

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files