Arca
Skip to content
Tools
Tools

Varian Unique Single Energy Linear Accelerator

Accelerator, Linear, Medical. 510(k) K101751, from Varian Medical Systems, Oncology Systems (Palo Alto, CA). Product code IYE, class 2.

As of Nov 20, 2026, K101751 is an FDA 510(k) held by Varian Medical Systems, Oncology Systems, substantially equivalent on Aug 10, 2010 after 49 days, with 1 recall tied to it in FDA's recall database.

  • Review time: 49 days
  • Recalls: 1
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The UNIQUE is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. *510(k) Summary UNIQUE - Varian Medical Systems* *Page 1 of 2*

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate
  • VenusXK232489 · Linatech, LLC · Apr 12, 2024
  • UniqueK132705 · Varian Medical Systems, Inc. · Oct 31, 2013

Recalls

Recalls FDA ties to K101751, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
StartedRecallStatus
Nov 28, 2011Z-0483-2012An event has been reported to Varian which entails excessive connector resistance, which caused the actual jaw positions to differ from the intended jaw positions without warning the operator.Terminated

Other IYE clearances

The newest 510(k)s for the same product code; their median review was 173 days; see every IYE device

Record

Type
Traditional
Received
Jun 22, 2010
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.