Arca
Skip to content
Tools
Tools

TX1 Tissue Removal System

Instrument, Ultrasonic Surgical. 510(k) K101561, from American Optisurgical, Inc. (Lake Forest, CA). Product code LFL, class U.

As of Nov 20, 2026, K101561 is an FDA 510(k) held by American Optisurgical, Inc., substantially equivalent on Aug 27, 2010 after 84 days, with 2 recalls tied to it in FDA's recall database.

  • Review time: 84 days
  • Recalls: 2
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

The TX1 Tissue Removal System is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft tissue are desirable, including General Surgery, Orthopedic Surgery, Laparoscopic Surgery and Plastic and Reconstructive Surgery.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K101561, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Mar 30, 2013Z-1516-2013The recall has been initiated due to an overheating protection voltage suppressor on the redundant safety protection circuit board. Secondary to the recall, American Optisurgical, Inc. will be making a correction to add…Terminated
Feb 22, 2012Z-1817-2012Two reasons for recall. 1. Incorrect pressure relief valve installed inside console, leading to a remote chance of a risk to health. 2. Technical Specifications, under Irrigation, the Operating Pressure value is…Terminated

Other LFL clearances

The newest 510(k)s for the same product code; their median review was 158 days; see every LFL device

Record

Type
Traditional
Received
Jun 4, 2010

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files

Esc

Type a name: a tool, a company, a drug or brand, a condition, a gene or a state. Identifiers like K231234, NCT05012345 or NDA 215866 open their record.

Logos are each company's own site icon, matched by name through Wikidata, or logos provided by Logo.dev.Company names and logos identify the companies in public FDA and registry records and do not imply endorsement.