Sedecal Mobile Digital Diagnostic X-Ray

System, X-Ray, Mobile. 510(k) K101517, from Sedecal., Sa. (Naples, FL). Product code IZL, class 2.

As of Nov 19, 2026, K101517 is an FDA 510(k) held by Sedecal., Sa., substantially equivalent on Aug 5, 2011 after 429 days, with 4 recalls tied to it in FDA's recall database.

  • Review time: 429 days
  • Recalls: 4
  • Cited as a predicate by: 2 devices

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

These Digital Radiographic Systems are intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

2 later 510(k) summaries name this device as a predicate

Recalls

Recalls FDA ties to K101517, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
Jun 10, 2025Z-0242-2026Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences…Open
Nov 2, 2018Z-0460-2019There is a potential for the RadPRO Mobile 40kW Digital X-Ray System Model SM-40HF-B-D-C to separate from the Tube Head Mount and potentially fall off the system.Terminated
Jul 13, 2015Z-2390-2015Electrostatic energy may be stored in Systems that can cause a short-cicuit which, in turn, can generate an Electomagnetic Pulse which might disturb the electronics of the Digital Motion Control Board producing…Terminated
Sep 22, 2014Z-1064-2015There is a potential for unintended and erratic motion.Terminated

Other IZL clearances

The newest 510(k)s for the same product code; their median review was 166 days; see every IZL device

Record

Type
Special
Received
Jun 2, 2010
Review panel
Radiology
Regulation
21 CFR 892.1720

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files