Arca
Skip to content
Tools
Tools

Modification To Iba Proton Therapy System-Proteus 235

System, Radiation Therapy, Charged-Particle, Medical. 510(k) K101508, from Ion Beam Applications S.A. (Philadelphia, PA). Product code LHN, class 2.

As of Nov 20, 2026, K101508 is an FDA 510(k) held by Ion Beam Applications S.A., substantially equivalent on Aug 6, 2010 after 66 days, with 6 recalls tied to it in FDA's recall database.

  • Review time: 66 days
  • Recalls: 6
  • Cited as a predicate by: 1 device

Indications for use

As the 510(k) summary states them, read from FDA's summary PDF

« The Proton Therapy System – Proteus 235 is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. »

Predicate devices

The legally marketed devices this 510(k) summary says it is substantially equivalent to, read from FDA's summary PDF

Cited as a predicate

1 later 510(k) summary names this device as a predicate

Recalls

Recalls FDA ties to K101508, newest first; recall numbers filed together for one problem share a row, and each opens FDA's recall record
May 26, 2021Z-2155-2021if the user forgets to select the trigger input manually while beam gating is prescribed for the patient, the proton irradiation will be delivered without gating. If the beam is not gated as prescribed, the treatment…Open
May 16, 2018Z-2424-2018Correction vector confirmation message is lost if access point is changed after sending the correction vector. As a result, the patient will be treated in the setup position or treatment position instead of the…Terminated
Jan 26, 2018Z-1477-2018When switching from Pencil Beam Scanning treatment mode to Uniform Scanning treatment mode, the tuning setpoints of the scanning magnets are not always taken into account when the proton beam is requested. The mismatch…Terminated
Nov 17, 2017Z-0938-2018There is a risk of misalignment of the patient when using treatment plans with multiple isocenters.Terminated
Oct 17, 2017Z-0930-2018IBA identified incorrect screw holes depth in graphite block 8 with respect to the degraderTerminated
Mar 9, 2017Z-1695-2017The Proton Therapy System can allow gantry rotation even if the snout is not properly locked.Terminated

Other LHN clearances

The newest 510(k)s for the same product code; their median review was 190 days; see every LHN device

Record

Type
Special
Received
Jun 1, 2010
Review panel
Radiology
Regulation
21 CFR 892.5050

One page per 510(k), De Novo and original PMA. Recalls are those FDA's recall database ties to this number; recalls of related devices that cite another number are not shown. Refreshed weekly. Predicates are read by Arca from the text of FDA's 510(k) summary PDFs, most of them scans: a device number counts when the summary names it as a predicate or as the device it is substantially equivalent to. About 97% of those read were right on a sample we checked by hand; summaries from before 2000 often name predicates without a number, so older devices show fewer. Pivotal studies are read by a model from FDA's De Novo decision summaries and PMA SSEDs; a count or result is shown only when its numbers appear in FDA's text. openFDA device 510(k), PMA and recall files